ABSTRACT
The situation for infants and young children requiring mechanical circulatory support for bridge to transplant or
recovery remains challenging. The under one year of age group has few options, although they are the largest
age group for pediatric transplantation and have the highest waitlist mortality. Although adult VADs are
sometimes used in larger children, the pump size and output mismatch are significant limitations. The Berlin
Heart EXCOR VAD is the only device approved for patients under approximately 6 years of age, but
thromboembolism and neurologic dysfunction occur at a rate 10 times higher than in adult VADs. The Jarvik
2015 Pediatric VAD is in early clinical testing but there are limitations to its use in infants and for right heart
support. The objective of this proposal is to complete the pre-clinical testing of the Penn State Infant
Ventricular Assist Device (VAD) which will be a safer VAD for left or right support. The Infant VAD has been
tested as a Left Ventricular Assist Device (LVAD) in 35 animal studies, and has demonstrated excellent
biocompatibility. Despite an aggressive protocol in which there has been little to no anticoagulation, there has
been no clinically evident thromboembolism or end organ dysfunction. The Penn State Infant VAD is a
pneumatic pulsatile device. Although pulsatile VADs have been replaced by continuous flow pumps in adults,
there are advantages to the pulsatile approach in infants. These include a low risk of thromboembolism and
pump thrombosis with low levels of anticoagulation, flexibility in cannulation for LVAD, RVAD or BiVAD
support, automatic preload–sensitive control of pump output, physiologic pulsatility, the absence of net zero or
negative blood flow through the VAD at low pump speed which occurs in rotary pumps, and absence of
acquired von Willebrand syndrome. We propose to complete the studies required for an IDE submission to the
FDA under the Early Feasibility Study program. The specific aims are:
Aim 1. Complete design controls and manufacturing process development for the VAD and cannulae,
including a new inlet cannula for the RVAD.
Aim 2. Complete the design of the portable pneumatic Driver to support an Early Feasibility Study.
Aim 3. Perform in vitro verification testing of the final pump and driver in the LVAD, RVAD, and BiVAD
configurations to demonstrate hemodynamic performance and acceptable hemolysis.
Aim 4. Demonstrate safety (acceptably low thromboembolism and normal operation) of the final pump
and driver in the LVAD and RVAD configurations in the high flow (3) and low flow ranges (3).
Aim 5. Complete the pre-IDE documentation, design the EFS clinical protocol, hold a pre-sub meeting
with FDA, and prepare an IDE submission.