Abstract
Two commercially available kits for determining blood ammonia were compared, an enzymatic method requiring a sample volume of 100μl and a nonenzymatic method requiring 1 ml of sample. Calibration curves for both kits were linear to 270 μmol/l, however correlation studies on plasma samples revealed a lack of agreement between methods (r = 0.825, Y = 0.76X + 51). Improved correlation occurs when portein interference is eliminated from the enzymatic procedure. Finally, the nonenzymatic method is modified to accommodate 100 μl of sample and compared to the original method.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 192-194 |
| Number of pages | 3 |
| Journal | Clinical Biochemistry |
| Volume | 9 |
| Issue number | C |
| DOIs | |
| State | Published - 1976 |