Abstract
A prototype neural probe designed for chronic measurements and recordings of localized temperature field in brain has been fabricated and tested. An array of through-film vanadium oxide thermistors has been deposited onto glass and flexible 20-μm polyimide substrates. High resistivity vanadium oxide was deposited using bias target ion-beam deposition. Functioning sensors were 10 μm × 10 μm × 85-nm thick. Tested samples on glass showed a mean thermistor temperature coefficient of resistance of -4.060 ± 0.004%/°C and nominal resistance R0 of 1.3 MΩ.
Original language | English (US) |
---|---|
Article number | 7378831 |
Pages (from-to) | 2211-2212 |
Number of pages | 2 |
Journal | IEEE Sensors Journal |
Volume | 16 |
Issue number | 8 |
DOIs | |
State | Published - Apr 15 2016 |
All Science Journal Classification (ASJC) codes
- Instrumentation
- Electrical and Electronic Engineering