Abstract
A procedure is developed for self-calibration of broadband, spherically focused ultrasonic transducers based on reciprocity. The input and received signals are measured in a pulse-echo configuration. These signals are used in conjunction with a multi-Gaussian beam model to obtain the electromechanical transfer function of the transducer. This calibration procedure is advantageous because it reduces the experimental configuration to a single transducer and a reflector. Experimental results indicate that the transfer function is insensitive to on-axis reflector placement. This result supports the feasibility of integrating the calibration procedure into actual testing in some situations.
Original language | English (US) |
---|---|
Pages (from-to) | EL236-EL241 |
Journal | Journal of the Acoustical Society of America |
Volume | 140 |
Issue number | 3 |
DOIs | |
State | Published - Sep 1 2016 |