Abstract
This work describes the apparatus for NIST-F4, an updated cesium atomic fountain at the National Institute of Standards and Technology (NIST), and presents an accuracy evaluation of the fountain as a primary frequency standard. The fountain uses optical molasses to laser cool a cloud of cesium atoms and launch it vertically in a fountain geometry. In high-density mode, the fractional frequency stability of NIST-F4 is σ y ( τ ) = 1.5 × 10 − 13 / τ , where τ is the measurement time in seconds. The short-term stability is limited by quantum projection noise and by phase noise from the local oscillator, an oven-controlled crystal oscillator operating at 5 MHz. Systematic frequency shifts and their uncertainties have been evaluated, resulting in a systematic (type B) fractional frequency uncertainty σ B = 2.2
| Original language | English (US) |
|---|---|
| Article number | 035002 |
| Journal | Metrologia |
| Volume | 62 |
| Issue number | 3 |
| DOIs | |
| State | Published - Jun 1 2025 |
All Science Journal Classification (ASJC) codes
- General Engineering
Fingerprint