Skip to main navigation Skip to search Skip to main content

Accuracy evaluation of primary frequency standard NIST-F4

  • Vladislav Gerginov
  • , Gregory W. Hoth
  • , Thomas P. Heavner
  • , Thomas E. Parker
  • , Kurt Gibble
  • , Jeff A. Sherman

Research output: Contribution to journalArticlepeer-review

Abstract

This work describes the apparatus for NIST-F4, an updated cesium atomic fountain at the National Institute of Standards and Technology (NIST), and presents an accuracy evaluation of the fountain as a primary frequency standard. The fountain uses optical molasses to laser cool a cloud of cesium atoms and launch it vertically in a fountain geometry. In high-density mode, the fractional frequency stability of NIST-F4 is σ y ( τ ) = 1.5 × 10 − 13 / τ , where τ is the measurement time in seconds. The short-term stability is limited by quantum projection noise and by phase noise from the local oscillator, an oven-controlled crystal oscillator operating at 5 MHz. Systematic frequency shifts and their uncertainties have been evaluated, resulting in a systematic (type B) fractional frequency uncertainty σ B = 2.2

Original languageEnglish (US)
Article number035002
JournalMetrologia
Volume62
Issue number3
DOIs
StatePublished - Jun 1 2025

All Science Journal Classification (ASJC) codes

  • General Engineering

Fingerprint

Dive into the research topics of 'Accuracy evaluation of primary frequency standard NIST-F4'. Together they form a unique fingerprint.