TY - JOUR
T1 - Adverse events associated with robotic-assistance in total hip arthroplasty
T2 - an analysis based on the FDA MAUDE database
AU - Graefe, S. Bradley
AU - Kirchner, Gregory J.
AU - Pahapill, Natalie K.
AU - Nam, Hannah H.
AU - Dunleavy, Mark L.
AU - Haines, Nikkole
N1 - Publisher Copyright:
© The Author(s) 2024.
PY - 2024/11
Y1 - 2024/11
N2 - Background: The Food and Drug Administration (FDA) maintains the Manufacturer and User Facility Device Experience (MAUDE) database for reporting adverse events associated with medical devices, including emerging technologies, such as robotic-assisted total hip arthroplasty (THA). Aim of this study was to evaluate the variation of adverse events associated with robotics in THA. Methods: Medical device reports (MDRs) within the MAUDE database were identified between 2017 and 2021. For MDR identification the product class “orthopaedic stereotaxic equipment” and terms associated with THA were used. Individual adverse events were identified and organised by type and consequences, such as patient injury, surgical delay, or conversion to the manual technique. Results: 521 MDRs constituting 546 discrete events were found. The most common reported complication was intraoperative hardware failure (304/546, 55.7%), among which the most common failure was a broken impaction handle/platform (110, 20.1%). Inaccurate cup placement was the second most common reported complication (63, 11.5%). Abandoning the robot occurred in 13.0% (71/521) of reports. A surgical delay was noted in 28% (146/521) of reports, with an average delay of 17.9 (range 1–60) minutes. Conclusions: Identifying complications that may occur with robotics in THA is an important first step in preventing adverse events and surgical delays. Database analysis provide an overview of the range of complications.
AB - Background: The Food and Drug Administration (FDA) maintains the Manufacturer and User Facility Device Experience (MAUDE) database for reporting adverse events associated with medical devices, including emerging technologies, such as robotic-assisted total hip arthroplasty (THA). Aim of this study was to evaluate the variation of adverse events associated with robotics in THA. Methods: Medical device reports (MDRs) within the MAUDE database were identified between 2017 and 2021. For MDR identification the product class “orthopaedic stereotaxic equipment” and terms associated with THA were used. Individual adverse events were identified and organised by type and consequences, such as patient injury, surgical delay, or conversion to the manual technique. Results: 521 MDRs constituting 546 discrete events were found. The most common reported complication was intraoperative hardware failure (304/546, 55.7%), among which the most common failure was a broken impaction handle/platform (110, 20.1%). Inaccurate cup placement was the second most common reported complication (63, 11.5%). Abandoning the robot occurred in 13.0% (71/521) of reports. A surgical delay was noted in 28% (146/521) of reports, with an average delay of 17.9 (range 1–60) minutes. Conclusions: Identifying complications that may occur with robotics in THA is an important first step in preventing adverse events and surgical delays. Database analysis provide an overview of the range of complications.
UR - http://www.scopus.com/inward/record.url?scp=85200457913&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85200457913&partnerID=8YFLogxK
U2 - 10.1177/11207000241263315
DO - 10.1177/11207000241263315
M3 - Article
C2 - 39099264
AN - SCOPUS:85200457913
SN - 1120-7000
VL - 34
SP - 688
EP - 694
JO - HIP International
JF - HIP International
IS - 6
ER -