Abstract
A preliminary examination of the secondary neutron dose from 18 MV Intensity Modulated Radiation Therapy (IMRT) was performed using the Monte Carlo code, MCNPX. A model of the Varian Clinac 2200C/2300C including the multileaf collimator was created to compute the neutron fluence and production in and around the treatment head. The neutron fluence results were folded with ICRP 74 fluence-to-dose conversion coefficients to obtain the ambient dose equivalent outside of the treatment head. The calculations are compared with measurements at the same location.
Original language | English (US) |
---|---|
Pages | 600-606 |
Number of pages | 7 |
State | Published - 2003 |
Event | Sixth International Meeting on Nuclear Applications of Accelerator Technology (AccApp'03): Accelerator Applications in a Nuclear Renaissance - San Diego, CA, United States Duration: Jun 1 2003 → Jun 5 2003 |
Other
Other | Sixth International Meeting on Nuclear Applications of Accelerator Technology (AccApp'03): Accelerator Applications in a Nuclear Renaissance |
---|---|
Country/Territory | United States |
City | San Diego, CA |
Period | 6/1/03 → 6/5/03 |
All Science Journal Classification (ASJC) codes
- General Engineering