Comparison of a new slender 6 Fr sheath with a standard 5 Fr sheath for transradial coronary angiography and intervention: RAP and BEAT (Radial Artery Patency and Bleeding, Efficacy, Adverse evenT), a randomised multicentre trial

Adel Aminian, Shigeru Saito, Akihiko Takahashi, Ivo Bernat, Robert Lee Jobe, Takashi Kajiya, Ian C. Gilchrist, Yves Louvard, Ferdinand Kiemeneij, Niels Van Royen, Seiji Yamazaki, Takashi Matsukage, Sunil V. Rao

Research output: Contribution to journalArticlepeer-review

60 Scopus citations

Abstract

Aims: The 6 Fr Glidesheath Slender (GSS6Fr) is a recently developed thin-walled radial sheath with an outer diameter (OD) that is smaller than the OD of standard 6 Fr sheaths. The purpose of this trial was to clarify whether the use of this new slender sheath would result in similar rates of RAO to a standard 5 Fr sheath in unselected patients undergoing transradial (TR) coronary angiography and/or intervention, and to assess the relative importance of sheath size and haemostasis protocol on the rate of RAO. Methods and results: We conducted a randomised, multicentre, non-inferiority trial comparing the GSS6Fr against the standard GS5Fr in patients undergoing TR coronary angiography and/or intervention. Patients in each group were subsequently randomised to undergo patent haemostasis or the institutional haemostasis protocol. The primary endpoint was the occurrence of RAO at discharge. A total of 1,926 patients were randomised in 12 centres. The incidence of RAO was 3.47% with GSS6Fr compared with 1.74% with GS5Fr (risk difference 1.73%, 95% CI: 0.51-2.95%; pnon-inferiority=0.150). Patients randomised to patent haemostasis had a similar rate of RAO compared with institutional haemostasis (2.61% vs. 2.61%, p=1). There was no difference with regard to all secondary endpoints, including vascular access-site complications, local bleeding and spasm. Conclusions: In this large multicentre randomised trial, the GSS6Fr was associated with a low event rate for the primary endpoint (RAO), although non-inferiority to the GS5Fr was not met, due to a lower than expected rate of RAO in the GS5Fr group. As compared to institutional haemostasis, the use of patent haemostasis was not associated with a reduced rate of RAO.

Original languageEnglish (US)
Pages (from-to)e549-e556
JournalEuroIntervention
Volume13
Issue number5
DOIs
StatePublished - Aug 2017

All Science Journal Classification (ASJC) codes

  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Comparison of a new slender 6 Fr sheath with a standard 5 Fr sheath for transradial coronary angiography and intervention: RAP and BEAT (Radial Artery Patency and Bleeding, Efficacy, Adverse evenT), a randomised multicentre trial'. Together they form a unique fingerprint.