T1 - Comparison of different inspiratory triggering settings in automated ventilators during cardiopulmonary resuscitation in a porcine model
AU - Tan, Dingyu
AU - Xu, Jun
AU - Shao, Shihuan
AU - Fu, Yangyang
AU - Sun, Feng
AU - Zhang, Yazhi
AU - Hu, Yingying
AU - Walline, Joseph
AU - Zhu, Huadong
AU - Yu, Xuezhong
N1 - Publisher Copyright:
© 2017 Tan et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2017/2
Y1 - 2017/2
N2 - Background Mechanical ventilation via automated in-hospital ventilators is quite common during cardiopulmonary resuscitation. It is not known whether different inspiratory triggering sensitivity settings of ordinary ventilators have different effects on actual ventilation, gas exchange and hemodynamics during resuscitation. Methods 18 pigs enrolled in this study were anaesthetized and intubated. Continuous chest compressions and mechanical ventilation (volume-controlled mode, 100% O2, respiratory rate 10/min, and tidal volumes 10ml/kg) were performed after 3 minutes of ventricular fibrillation. Group trig-4, trig-10 and trig-20 (six pigs each) were characterized by triggering sensitivities of 4, 10 and 20 (cmH2O for pressure-triggering and L/min for flow-triggering), respectively. Additionally, each pig in each group was mechanically ventilated using three types of inspiratory triggering (pressure-triggering, flow-triggering and turned-off triggering) of 5 minutes duration each, and each animal matched with one of six random assortments of the three different triggering settings. Blood gas samples, respiratory and hemodynamic parameters for each period were all collected and analyzed. Results In each group, significantly lower actual respiratory rate, minute ventilation volume, mean airway pressure, arterial pH, PaO2, and higher end-tidal carbon dioxide, aortic blood pressure, coronary perfusion pressure, PaCO2 and venous oxygen saturation were observed in the ventilation periods with a turned-off triggering setting compared to those with pressureor flow- triggering (all P<0.05), except when compared with pressure-triggering of 20 cmH2O (respiratory rate 10.5[10/11.3]/min vs 12.5[10.8/13.3]/min, P = 0.07; coronary perfusion pressure 30.3[24.5/31.6] mmHg vs 27.4[23.7/29] mmHg, P = 0.173; venous oxygen saturation 46.5[32/56.8]% vs 41.5[33.5/48.5]%, P = 0.575). Conclusions Ventilation with pressure- or flow-triggering tends to induce hyperventilation and deteriorating gas exchange and hemodynamics during CPR. A turned-off patient triggering or a pressure- triggering of 20 cmH2O is preferred for ventilation when an ordinary inpatient hospital ventilator is used during resuscitation.
AB - Background Mechanical ventilation via automated in-hospital ventilators is quite common during cardiopulmonary resuscitation. It is not known whether different inspiratory triggering sensitivity settings of ordinary ventilators have different effects on actual ventilation, gas exchange and hemodynamics during resuscitation. Methods 18 pigs enrolled in this study were anaesthetized and intubated. Continuous chest compressions and mechanical ventilation (volume-controlled mode, 100% O2, respiratory rate 10/min, and tidal volumes 10ml/kg) were performed after 3 minutes of ventricular fibrillation. Group trig-4, trig-10 and trig-20 (six pigs each) were characterized by triggering sensitivities of 4, 10 and 20 (cmH2O for pressure-triggering and L/min for flow-triggering), respectively. Additionally, each pig in each group was mechanically ventilated using three types of inspiratory triggering (pressure-triggering, flow-triggering and turned-off triggering) of 5 minutes duration each, and each animal matched with one of six random assortments of the three different triggering settings. Blood gas samples, respiratory and hemodynamic parameters for each period were all collected and analyzed. Results In each group, significantly lower actual respiratory rate, minute ventilation volume, mean airway pressure, arterial pH, PaO2, and higher end-tidal carbon dioxide, aortic blood pressure, coronary perfusion pressure, PaCO2 and venous oxygen saturation were observed in the ventilation periods with a turned-off triggering setting compared to those with pressureor flow- triggering (all P<0.05), except when compared with pressure-triggering of 20 cmH2O (respiratory rate 10.5[10/11.3]/min vs 12.5[10.8/13.3]/min, P = 0.07; coronary perfusion pressure 30.3[24.5/31.6] mmHg vs 27.4[23.7/29] mmHg, P = 0.173; venous oxygen saturation 46.5[32/56.8]% vs 41.5[33.5/48.5]%, P = 0.575). Conclusions Ventilation with pressure- or flow-triggering tends to induce hyperventilation and deteriorating gas exchange and hemodynamics during CPR. A turned-off patient triggering or a pressure- triggering of 20 cmH2O is preferred for ventilation when an ordinary inpatient hospital ventilator is used during resuscitation.
UR - http://www.scopus.com/inward/record.url?scp=85012305776&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85012305776&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0171869
DO - 10.1371/journal.pone.0171869
M3 - Article
C2 - 28187154
AN - SCOPUS:85012305776
SN - 1932-6203
VL - 12
JO - PloS one
JF - PloS one
IS - 2
M1 - e0171869
ER -