TY - JOUR
T1 - Comparison of two types of neonatal extracorporeal life support systems with pulsatile and nonpulsatile flow
AU - Haines, Nikkole
AU - Wang, Shigang
AU - Myers, John L.
AU - Ündar, Akif
PY - 2009/11
Y1 - 2009/11
N2 - We compared the effects of two neonatal extracorporeal life support (ECLS) systems on circuit pressures and surplus hemodynamic energy levels in a simulated ECLS model. The clinical set-up included the Jostra HL-20 heart-lung machine, either the Medtronic ECMO (0800) or the MEDOS 800LT systems with company-provided circuit components, a 10 Fr arterial cannula, and a pseudo-patient. We tested the system in nonpulsatile and pulsatile flow modes at two flow rates using a 40/60 glycerin/water blood analog, for a total of 48 trials, with n = 6 for each set-up. The pressure drops over the Medtronic ECLS were significantly higher than those over the MEDOS system regardless of the flow rate or perfusion mode (144.8 ± 0.2 mm Hg vs. 35.7 ± 0.2 mm Hg, respectively, at 500 mL/min in nonpulsatile mode, P < 0.001). The preoxygenator mean arterial pressures were significantly increased and the precannula hemodynamic energy values were decreased with the Medtronic ECLS circuit. These results suggest that the MEDOS ECLS circuit better transmits hemodynamic energy to the patient, keeps mean circuit pressures lower, and has lower pressure drops than the Medtronic Circuit.
AB - We compared the effects of two neonatal extracorporeal life support (ECLS) systems on circuit pressures and surplus hemodynamic energy levels in a simulated ECLS model. The clinical set-up included the Jostra HL-20 heart-lung machine, either the Medtronic ECMO (0800) or the MEDOS 800LT systems with company-provided circuit components, a 10 Fr arterial cannula, and a pseudo-patient. We tested the system in nonpulsatile and pulsatile flow modes at two flow rates using a 40/60 glycerin/water blood analog, for a total of 48 trials, with n = 6 for each set-up. The pressure drops over the Medtronic ECLS were significantly higher than those over the MEDOS system regardless of the flow rate or perfusion mode (144.8 ± 0.2 mm Hg vs. 35.7 ± 0.2 mm Hg, respectively, at 500 mL/min in nonpulsatile mode, P < 0.001). The preoxygenator mean arterial pressures were significantly increased and the precannula hemodynamic energy values were decreased with the Medtronic ECLS circuit. These results suggest that the MEDOS ECLS circuit better transmits hemodynamic energy to the patient, keeps mean circuit pressures lower, and has lower pressure drops than the Medtronic Circuit.
UR - http://www.scopus.com/inward/record.url?scp=70450278669&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=70450278669&partnerID=8YFLogxK
U2 - 10.1111/j.1525-1594.2009.00934.x
DO - 10.1111/j.1525-1594.2009.00934.x
M3 - Article
C2 - 19874284
AN - SCOPUS:70450278669
SN - 0160-564X
VL - 33
SP - 958
EP - 966
JO - Artificial organs
JF - Artificial organs
IS - 11
ER -