Abstract

A small-scale piezoelectric actuator has been designed for use in minimally invasive surgery. The actuator is designed to actuate a compliant suture needle grasper as part of a telerobotic surgical system. The actuator consists of a single piezoceramic stack and a slider-crank mechanism which acts as a mechanical stroke amplifier. The main focus of this paper is on the design of the actuator and fabrication and test of a prototype as a proof of concept. The design and fabrication of the prototype actuator is discussed as well as experimental results measuring its performance. A quasi-static model of the actuator which predicts the mechanical amplification and force response is also discussed. Results from laboratory bench-top experiments are presented, and the actuator together with the compliant gripper are shown to grasp and hold a standard suture needle.

Original languageEnglish (US)
Pages (from-to)299-305
Number of pages7
JournalAmerican Society of Mechanical Engineers, Aerospace Division (Publication) AD
Volume60
StatePublished - 2000

All Science Journal Classification (ASJC) codes

  • Mechanical Engineering
  • Space and Planetary Science

Fingerprint

Dive into the research topics of 'Design and modeling of a piezoelectric actuator for minimally invasive surgery'. Together they form a unique fingerprint.