Abstract
A small-scale piezoelectric actuator has been designed for use in minimally invasive surgery. The actuator is designed to actuate a compliant suture needle grasper as part of a telerobotic surgical system. The actuator consists of a single piezoceramic stack and a slider-crank mechanism which acts as a mechanical stroke amplifier. The main focus of this paper is on the design of the actuator and fabrication and test of a prototype as a proof of concept. The design and fabrication of the prototype actuator is discussed as well as experimental results measuring its performance. A quasi-static model of the actuator which predicts the mechanical amplification and force response is also discussed. Results from laboratory bench-top experiments are presented, and the actuator together with the compliant gripper are shown to grasp and hold a standard suture needle.
Original language | English (US) |
---|---|
Pages (from-to) | 299-305 |
Number of pages | 7 |
Journal | American Society of Mechanical Engineers, Aerospace Division (Publication) AD |
Volume | 60 |
State | Published - 2000 |
All Science Journal Classification (ASJC) codes
- Mechanical Engineering
- Space and Planetary Science