EJ-309 pulse shape discrimination performance with a high gamma-ray-to-neutron ratio and low threshold

A. C. Kaplan, M. Flaska, A. Enqvist, J. L. Dolan, S. A. Pozzi

Research output: Contribution to journalArticlepeer-review

29 Scopus citations

Abstract

Measuring neutrons in the presence of high gamma-ray fluence is a challenge with multi-particle detectors. Organic liquid scintillators such as the EJ-309 are capable of accurate pulse-shape discrimination (PSD) but the chance for particle misclassification is not negligible for some applications. By varying the distance from an EJ-309 scintillator to a strong-gamma-ray source and keeping a weak-neutron source at a fixed position, various gamma-to-neutron ratios can be measured and PSD performance can be quantified. Comparing neutron pulse-height distributions allows for pulse-height specific PSD evaluation, and quantification and visualization of deviation from 252Cf alone. Even with the addition of the misclassified gamma-rays, the PSD is effective in separating particles so that neutron count rate can be predicted with less than 10% error up to a gamma-to-neutron ratio of almost 650. For applications which can afford a reduction in neutron detection efficiency, PSD can be sufficiently effective in discriminating particles to measure a weak neutron source in a high gamma-ray background.

All Science Journal Classification (ASJC) codes

  • Nuclear and High Energy Physics
  • Instrumentation

Fingerprint

Dive into the research topics of 'EJ-309 pulse shape discrimination performance with a high gamma-ray-to-neutron ratio and low threshold'. Together they form a unique fingerprint.