T1 - Expansion anchors for use in anterior cruciate ligament (ACL) reconstruction
T2 - Establishing proof of concept in a benchtop analysis
AU - Black, K. P.
AU - Saunders, M. M.
N1 - Funding Information:
The authors gratefully acknowledge the contributions of Don Whitehaus of Whitehaus Precision Machine for fabrication of the anchor systems evaluated in this study. This work was funded by The Department of Orthopaedics and Rehabilitation, The Pennsylvania State University College of Medicine.
PY - 2005/6
Y1 - 2005/6
N2 - The current method for graft fixation in bone tendon-bone anterior cruciate ligament (ACL) reconstruction is the interference screw. Although this method of fixation provides for adequate graft fixation with respect to strength, intraoperative placement is difficult and the failure rate is high. To address these concerns, we have designed and fabricated prototype expansion anchors that could be expanded to anchor the graft in the bone tunnel. As a first step in assessing the validity of this concept, in the current work, we demonstrate that these systems are of comparable fixation strength (biomechanical pullout testing) to the standard interference screw, are smaller at the time of insertion and thus provide for increased visibility and ease of placement. The increased visibility should result in better placement and reduced failure rates. The increased ease of placement should result in significant savings in decreased OR time.
AB - The current method for graft fixation in bone tendon-bone anterior cruciate ligament (ACL) reconstruction is the interference screw. Although this method of fixation provides for adequate graft fixation with respect to strength, intraoperative placement is difficult and the failure rate is high. To address these concerns, we have designed and fabricated prototype expansion anchors that could be expanded to anchor the graft in the bone tunnel. As a first step in assessing the validity of this concept, in the current work, we demonstrate that these systems are of comparable fixation strength (biomechanical pullout testing) to the standard interference screw, are smaller at the time of insertion and thus provide for increased visibility and ease of placement. The increased visibility should result in better placement and reduced failure rates. The increased ease of placement should result in significant savings in decreased OR time.
UR - http://www.scopus.com/inward/record.url?scp=18044386535&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=18044386535&partnerID=8YFLogxK
U2 - 10.1016/j.medengphy.2004.11.004
DO - 10.1016/j.medengphy.2004.11.004
M3 - Article
C2 - 15863351
AN - SCOPUS:18044386535
SN - 1350-4533
VL - 27
SP - 425
EP - 434
JO - Medical Engineering and Physics
JF - Medical Engineering and Physics
IS - 5
ER -