Abstract
We report the first accuracy evaluation of NRC-FCs2, an atomic fountain clock developed at the National Research Council Canada. The short term stability at high atomic density is 1.1 10-13 at 1 second of averaging time. The typical overall fractional type B uncertainty is 2.3 10-16. GNSS-based frequency comparisons of NRC-FCs2 with the SI second, as reported by the BIPM over fourteen months, give a fractional frequency difference of -2.4(3.7) 10-16
Original language | English (US) |
---|---|
Article number | 035010 |
Journal | Metrologia |
Volume | 57 |
Issue number | 3 |
DOIs | |
State | Published - Jun 2020 |
All Science Journal Classification (ASJC) codes
- Engineering(all)