TY - JOUR
T1 - Flow visualization of a pediatric ventricular assist device during stroke volume reductions related to weaning
AU - Roszelle, Breigh N.
AU - Deutsch, Steven
AU - Weiss, William J.
AU - Manning, Keefe B.
N1 - Funding Information:
This study was supported by NIH—NHLBI Contract HV 48191 and the ARL Exploratory and Foundational Graduate Student Research Program.
PY - 2011/7
Y1 - 2011/7
N2 - The aim of this study is to define the fluid mechanics of a pulsatile pneumatically driven pediatric ventricular assist device (PVAD), for the reduced flow rates encountered during device weaning and myocardial recovery, and relate the results to the potential for thromboembolic events. We place an acrylic model of the PVAD in a mock circulatory loop filled with a viscoelastic blood analog and operate at four stroke volumes (SVs), each with two different filling conditions, to mimic how the flow rate of the device may be reduced. Particle image velocimetry is used to acquire flow field data. We find that a SV reduction method provides better rotational flow and higher wall shear rates than a beat rate reduction method; that a quick filling condition with a compressed diastolic time is better than a slow filling condition; and, that a reduction in SV to 40% led to greatly reduced fluid movement and wall shear rates that could increase the thrombogenicity of the device. SV reduction is a viable option for flow reduction during weaning, however, it does lead to significant changes to the device flow field and future studies are needed to develop operational protocols for the PVAD during bridge-to-recovery.
AB - The aim of this study is to define the fluid mechanics of a pulsatile pneumatically driven pediatric ventricular assist device (PVAD), for the reduced flow rates encountered during device weaning and myocardial recovery, and relate the results to the potential for thromboembolic events. We place an acrylic model of the PVAD in a mock circulatory loop filled with a viscoelastic blood analog and operate at four stroke volumes (SVs), each with two different filling conditions, to mimic how the flow rate of the device may be reduced. Particle image velocimetry is used to acquire flow field data. We find that a SV reduction method provides better rotational flow and higher wall shear rates than a beat rate reduction method; that a quick filling condition with a compressed diastolic time is better than a slow filling condition; and, that a reduction in SV to 40% led to greatly reduced fluid movement and wall shear rates that could increase the thrombogenicity of the device. SV reduction is a viable option for flow reduction during weaning, however, it does lead to significant changes to the device flow field and future studies are needed to develop operational protocols for the PVAD during bridge-to-recovery.
UR - http://www.scopus.com/inward/record.url?scp=79958844994&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=79958844994&partnerID=8YFLogxK
U2 - 10.1007/s10439-011-0291-8
DO - 10.1007/s10439-011-0291-8
M3 - Article
C2 - 21404124
AN - SCOPUS:79958844994
SN - 0090-6964
VL - 39
SP - 2046
EP - 2058
JO - Annals of Biomedical Engineering
JF - Annals of Biomedical Engineering
IS - 7
ER -