TY - JOUR
T1 - Flow visualization of three-dimensionality inside the 12 cc penn state pulsatile pediatric ventricular assist device
AU - Roszelle, Breigh N.
AU - Deutsch, Steven
AU - Manning, Keefe B.
N1 - Funding Information:
Supported by NIH NHLBI Grant HV 48191.
PY - 2010/2
Y1 - 2010/2
N2 - In order to aid the ongoing concern of limited organ availability for pediatric heart transplants, Penn State has continued development of a pulsatile Pediatric Ventricular Assist Device (PVAD). Initial studies of the PVAD observed an increase in thrombus formation due to differences in flow field physics when compared to adult sized devices, which included a higher degree of three-dimensionality. This unique flow field brings into question the use of 2D planar particle image velocimetry (PIV) as a flow visualization technique, however the small size and high curvature of the PVAD make other tools such as stereoscopic PIV impractical. In order to test the reliability of the 2D results, we perform a pseudo-3D PIV study using planes both parallel and normal to the diaphragm employing a mock circulatory loop containing a viscoelastic fluid that mimics 40% hematocrit blood. We find that while the third component of velocity is extremely helpful to a physical understanding of the flow, particularly of the diastolic jet and the development of a desired rotational pattern, the flow data taken parallel to the diaphragm is sufficient to describe the wall shear rates, a critical aspect to the study of thrombosis and design of such pumps.
AB - In order to aid the ongoing concern of limited organ availability for pediatric heart transplants, Penn State has continued development of a pulsatile Pediatric Ventricular Assist Device (PVAD). Initial studies of the PVAD observed an increase in thrombus formation due to differences in flow field physics when compared to adult sized devices, which included a higher degree of three-dimensionality. This unique flow field brings into question the use of 2D planar particle image velocimetry (PIV) as a flow visualization technique, however the small size and high curvature of the PVAD make other tools such as stereoscopic PIV impractical. In order to test the reliability of the 2D results, we perform a pseudo-3D PIV study using planes both parallel and normal to the diaphragm employing a mock circulatory loop containing a viscoelastic fluid that mimics 40% hematocrit blood. We find that while the third component of velocity is extremely helpful to a physical understanding of the flow, particularly of the diastolic jet and the development of a desired rotational pattern, the flow data taken parallel to the diaphragm is sufficient to describe the wall shear rates, a critical aspect to the study of thrombosis and design of such pumps.
UR - http://www.scopus.com/inward/record.url?scp=77249101490&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77249101490&partnerID=8YFLogxK
U2 - 10.1007/s10439-009-9842-7
DO - 10.1007/s10439-009-9842-7
M3 - Article
C2 - 19936926
AN - SCOPUS:77249101490
SN - 0090-6964
VL - 38
SP - 439
EP - 455
JO - Annals of Biomedical Engineering
JF - Annals of Biomedical Engineering
IS - 2
ER -