TY - JOUR
T1 - Hematocrit determination in small bore tubes by differential spectrophotometry
AU - Lipowsky, Herbert H.
AU - Usami, Shunichi
AU - Chien, Shu
AU - Pittman, Roland N.
N1 - Funding Information:
Lipowsky. 3 Supported by USPHS Research Grants HL18292 and HL26901, and Am-Heart Association Grant-in-Aid 78915.
Funding Information:
’ To whom correspondence should be addressed. * Supported by USPHS Research Grants HLl6851, HL25305, and RCDA HL00594 for Dr.
PY - 1982/7
Y1 - 1982/7
N2 - A method for the measurement of the hematocrit of blood flowing in small-bore glass tubes (HT) is examined which relies upon a linear relationship between the difference in optical densities (ΔOD) at two isobestic wavelengths for oxy- and deoxyhemoglobin (546 and 520 nm) and the product of HT and luminal diameter (DT). The foundation for the technique has been shown to be firmly based upon the principles that: (1) The total optical density of blood flowing in a smallbore glass tube may be expressed as the sum of a pure absorption term (ODabs), which is linear with HT, and a nonlinear scattering term (B); and (2) at two closely spaced wavelengths, the scattering terms are approximately equal such that subtraction of the two total optical densities results in a term linearly proportional to the product of the difference in molecular extinction coefficients (Δε), hemoglobin concentration, and path length, DT. These relationships were examined for the case of blood flowing in tubes ranging from 40 to 70 μm in luminal diameter and yielded a linear relationship of ΔOD = (ε546 - ε520)ccell × HTDT, where ccell is the mean corpuscular hemoglobin concentration. The slopes of these calibration curves, obtained for hemoglobin solutions and red cell suspensions, agreed with macrospectrophotometric values of Δε to within 2 and 12%, respectively. Simulation of a blanket application of the method to the measurement of HT in these tubes revealed an accuracy of 11.6% (rms error) between values of HT computed from the regression curve and those obtained by centrifugation. It is also shown that by choosing isobestic wavelengths with extinction coefficients in the ratio of 2:1, the scattering term may be computed from the relationship B = OD520 - ΔOD; and that variations in B attendant to red cell shape changes, as for example in the case of the sickling of Hb SS cells during deoxygenation, may be easily evaluated.
AB - A method for the measurement of the hematocrit of blood flowing in small-bore glass tubes (HT) is examined which relies upon a linear relationship between the difference in optical densities (ΔOD) at two isobestic wavelengths for oxy- and deoxyhemoglobin (546 and 520 nm) and the product of HT and luminal diameter (DT). The foundation for the technique has been shown to be firmly based upon the principles that: (1) The total optical density of blood flowing in a smallbore glass tube may be expressed as the sum of a pure absorption term (ODabs), which is linear with HT, and a nonlinear scattering term (B); and (2) at two closely spaced wavelengths, the scattering terms are approximately equal such that subtraction of the two total optical densities results in a term linearly proportional to the product of the difference in molecular extinction coefficients (Δε), hemoglobin concentration, and path length, DT. These relationships were examined for the case of blood flowing in tubes ranging from 40 to 70 μm in luminal diameter and yielded a linear relationship of ΔOD = (ε546 - ε520)ccell × HTDT, where ccell is the mean corpuscular hemoglobin concentration. The slopes of these calibration curves, obtained for hemoglobin solutions and red cell suspensions, agreed with macrospectrophotometric values of Δε to within 2 and 12%, respectively. Simulation of a blanket application of the method to the measurement of HT in these tubes revealed an accuracy of 11.6% (rms error) between values of HT computed from the regression curve and those obtained by centrifugation. It is also shown that by choosing isobestic wavelengths with extinction coefficients in the ratio of 2:1, the scattering term may be computed from the relationship B = OD520 - ΔOD; and that variations in B attendant to red cell shape changes, as for example in the case of the sickling of Hb SS cells during deoxygenation, may be easily evaluated.
UR - http://www.scopus.com/inward/record.url?scp=0019955258&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0019955258&partnerID=8YFLogxK
U2 - 10.1016/0026-2862(82)90041-3
DO - 10.1016/0026-2862(82)90041-3
M3 - Article
C2 - 7121311
AN - SCOPUS:0019955258
SN - 0026-2862
VL - 24
SP - 42
EP - 55
JO - Microvascular Research
JF - Microvascular Research
IS - 1
ER -