Abstract
A study has been conducted with the goal of comparing the characteristics of TLD-K detectors based on sodium silicate ceramics to the characteristics of monocrystalline lithium fluoride detectors that contain traces of titanium and magnesium (TLD-100), as well as anion-defected aluminum oxide (TLD-500), which are widely used in thermoluminescent dosimetry. Because they are soil-equivalent, SiO2 detectors are well suited for measuring the absorbed radiation dose in soils. The results of the comparison indicate that TLD-K detectors are preferable to LiF detectors in environmental applications due to the better uniformity of the detector sensitivity in a batch, the wider range of measurable doses, and the lower threshold in determining small doses.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 733-739 |
| Number of pages | 7 |
| Journal | Instruments and Experimental Techniques |
| Volume | 59 |
| Issue number | 5 |
| DOIs | |
| State | Published - Sep 1 2016 |
All Science Journal Classification (ASJC) codes
- Instrumentation