T1 - Hydrodynamic comparison of the Penumbra system and commonly available syringes in forced-suction thrombectomy
AU - Simon, Scott Douglas
AU - Grey, Casey Paul
PY - 2014/4
Y1 - 2014/4
N2 - Background: The Penumbra system uses a coaxial separator and continuous extracorporeal suction to remove a clot from a cerebral artery. Forced-suction thrombectomy (FST) involves aspirating clots through the same reperfusion catheter using only a syringe, decreasing the procedure time and supplies needed. Objective: To evaluate multiple combinations of catheters and syringes to determine the optimal pairing for use in FST. Methods: Tests were performed using both the Penumbra system and syringes to aspirate water through Penumbra 0.041 inch (041), 4Max, 0.054 inch (054) and 5Max reperfusion catheters and a shuttle sheath. Dynamic pressure and flow at the catheter tip were calculated from the fill times for each system. Static pressure and force for each aspiration source were determined with a vacuum gauge. Results: All syringes provided significantly higher dynamic pressure at the catheter tip than the Penumbra system (p<0.001). Increasing syringe volume significantly increased static pressure (p<0.001). Both flow and aspiration force significantly increased with catheter size (p<0.001). Cases are presented to demonstrate the clinical value of the laboratory principles. Conclusions: Maximizing static and dynamic pressure when performing FST is achieved by aspirating with a syringe possessing both the largest volume and the largest inlet diameter available. Maximizing aspiration force and flow rate is achieved by using the largest catheter possible.
AB - Background: The Penumbra system uses a coaxial separator and continuous extracorporeal suction to remove a clot from a cerebral artery. Forced-suction thrombectomy (FST) involves aspirating clots through the same reperfusion catheter using only a syringe, decreasing the procedure time and supplies needed. Objective: To evaluate multiple combinations of catheters and syringes to determine the optimal pairing for use in FST. Methods: Tests were performed using both the Penumbra system and syringes to aspirate water through Penumbra 0.041 inch (041), 4Max, 0.054 inch (054) and 5Max reperfusion catheters and a shuttle sheath. Dynamic pressure and flow at the catheter tip were calculated from the fill times for each system. Static pressure and force for each aspiration source were determined with a vacuum gauge. Results: All syringes provided significantly higher dynamic pressure at the catheter tip than the Penumbra system (p<0.001). Increasing syringe volume significantly increased static pressure (p<0.001). Both flow and aspiration force significantly increased with catheter size (p<0.001). Cases are presented to demonstrate the clinical value of the laboratory principles. Conclusions: Maximizing static and dynamic pressure when performing FST is achieved by aspirating with a syringe possessing both the largest volume and the largest inlet diameter available. Maximizing aspiration force and flow rate is achieved by using the largest catheter possible.
UR - http://www.scopus.com/inward/record.url?scp=84897954085&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84897954085&partnerID=8YFLogxK
U2 - 10.1136/neurintsurg-2012-010638
DO - 10.1136/neurintsurg-2012-010638
M3 - Article
C2 - 23531712
AN - SCOPUS:84897954085
SN - 1759-8478
VL - 6
SP - 205
EP - 211
JO - Journal of neurointerventional surgery
JF - Journal of neurointerventional surgery
IS - 3
ER -