Interlaboratory comparison of 25-hydroxyvitamin D assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 — Part 1 liquid chromatography – tandem mass spectrometry (LC-MS/MS) assays — impact of 3-epi-25-hydroxyvitamin D3 on assay performance

  • Stephen A. Wise
  • , Johanna E. Camara
  • , Carolyn Q. Burdette
  • , Grace Hahm
  • , Federica Nalin
  • , Adam J. Kuszak
  • , Joyce Merkel
  • , Ramón A. Durazo-Arvizu
  • , Emma L. Williams
  • , Andrew N. Hoofnagle
  • , Fiona Ivison
  • , Ralf Fischer
  • , Jody M.W. van den Ouweland
  • , Chung S. Ho
  • , Emmett W.K. Law
  • , Jean Nicolas Simard
  • , Renaud Gonthier
  • , Brett Holmquist
  • , Sarah Meadows
  • , Lorna Cox
  • Kimberly Robyak, Michael H. Creer, Robert Fitzgerald, Michael W. Clarke, Norma Breen, Pierre Lukas, Étienne Cavalier, Christopher T. Sempos

Research output: Contribution to journalArticlepeer-review

Abstract

An interlaboratory comparison study was conducted by the Vitamin D Standardization Program (VDSP) to assess the performance of liquid chromatography – tandem mass spectrometry (LC-MS/MS) assays used for the determination of serum total 25-hydroxyvitamin D (25(OH)D), which is the sum of 25-hydroxyvitamin D2 (25(OH)D2) and 25-hydroxyvitamin D3 (25(OH)D3). A set of 50 single-donor samples was assigned target values for concentrations of 25(OH)D2, 25(OH)D3, 3-epi-25-hydroxyvitamin D3 (3-epi-25(OH)D3), and 24R,25-dihydroxyvitamin D3 (24R,25(OH)2D3) using isotope dilution liquid chromatography – tandem mass spectrometry (ID LC-MS/MS). VDSP Intercomparison Study 2 Part 1 includes results from 14 laboratories using 14 custom LC-MS/MS assays. Assay performance was evaluated using mean % bias compared to the assigned target values and using linear regression analysis of the test assay mean results and the target values. Only 53% of the LC-MS/MS assays met the VDSP criterion of mean % bias ≤ |±5%|. For the LC-MS/MS assays not meeting the ≤ |±5%| criterion, four assays had mean % bias of between 12 and 21%. Based on multivariable regression analysis using the concentrations of the four individual vitamin D metabolites in the 50 single-donor samples, the performance of several LC-MS/MS assays was found to be influenced by the presence of 3-epi-25(OH)D3. The results of this interlaboratory study represent the most comprehensive comparison of LC-MS/MS assay performance for serum total 25(OH)D and document the significant impact of the lack of separation of 3-epi-25(OH)D3 and 25(OH)D3 on assay performance, particularly with regard to mean % bias. Graphical abstract: [Figure not available: see fulltext.].

Original languageEnglish (US)
Pages (from-to)333-349
Number of pages17
JournalAnalytical and Bioanalytical Chemistry
Volume414
Issue number1
DOIs
StatePublished - Jan 2022

All Science Journal Classification (ASJC) codes

  • Analytical Chemistry
  • Biochemistry

Fingerprint

Dive into the research topics of 'Interlaboratory comparison of 25-hydroxyvitamin D assays: Vitamin D Standardization Program (VDSP) Intercomparison Study 2 — Part 1 liquid chromatography – tandem mass spectrometry (LC-MS/MS) assays — impact of 3-epi-25-hydroxyvitamin D3 on assay performance'. Together they form a unique fingerprint.

Cite this