Limitations of current validation protocols for home blood pressure monitors for individual patients

William Gerin, Amy R. Schwartz, Joseph E. Schwartz, Thomas G. Pickering, Karina W. Davidson, Jonathan Bress, Eoin O'Brien, Neil Atkins

Research output: Contribution to journalArticlepeer-review

78 Scopus citations

Abstract

Background: Automatic blood pressure monitoring conducted at home is increasingly used in the diagnosis and management of hypertension. We assessed the adequacy of existing British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) validation standards for automatic blood pressure monitoring devices. Subject and methods: A theoretical study and an empirical test are presented to estimate the proportion of persons for whom a blood pressure monitor validated according to existing BHS and AAMI standards would be inaccurate. Results: The results suggest that a major limitation of both protocols is the lack of attention given to the number of individual patients for whom a monitor may be inaccurate. A blood pressure monitor that meets the AAMI and BHS validation criteria may report blood pressures in error by more than 5 mmHg for more than half of the people. Conclusions: A validation standard that does not take account of the person-effects on error will lead to a substantial proportion of persons using self-monitors that are systematically inaccurate for that person.

Original languageEnglish (US)
Pages (from-to)313-318
Number of pages6
JournalBlood Pressure Monitoring
Volume7
Issue number6
DOIs
StatePublished - Dec 2002

All Science Journal Classification (ASJC) codes

  • Internal Medicine
  • Cardiology and Cardiovascular Medicine
  • Assessment and Diagnosis
  • Advanced and Specialized Nursing

Fingerprint

Dive into the research topics of 'Limitations of current validation protocols for home blood pressure monitors for individual patients'. Together they form a unique fingerprint.

Cite this