Abstract
Jointly developed by members of the American Society for Artificial Internal Organs and the Society of Thoracic Surgeons along with staff from the Food and Drug Administration, the National Heart, Lung and Blood Institute and other experts, this recommendation describes the reliability considerations and goals for Investigational Device Exemption and Premarket Approval submissions for long-term, mechanical circulatory support systems. The recommendation includes a definition of system failure, a discussion of an appropriate reliability model, a suggested in vitro reliability test plan, reliability considerations for animal implantation tests, in vitro and animal in vivo performance goals, the qualification of design changes during the Investigational Device Exemption clinical trial, the development of a Failure Modes Effects and Criticality Analysis, and the reliability information for surgeons and patient candidates. The document will be periodically reviewed to assess its timeliness and appropriateness within five years.
Original language | English (US) |
---|---|
Pages (from-to) | 1852-1859 |
Number of pages | 8 |
Journal | Annals of Thoracic Surgery |
Volume | 66 |
Issue number | 5 |
DOIs | |
State | Published - 1998 |
All Science Journal Classification (ASJC) codes
- Surgery
- Pulmonary and Respiratory Medicine
- Cardiology and Cardiovascular Medicine