MICRO-CALORIMETER FOR MONOLAYER FILMS.

J. H. Campbell, M. Bretz, M. H.W. Chan

Research output: Chapter in Book/Report/Conference proceedingConference contribution

1 Scopus citations

Abstract

Adsorbed monolayer specific heat experiments have relied to date on exfoliated substrates with large sample areas but complicated geometry. Using micro-circuit fabrication techniques, we have instead constructed an ac calorimeter from a small cleaved sheet of graphite. Total heat capacity at 3 K is 45 multiplied by 10** minus **9 J/K, with a film contribution of approximately 1 multiplied by 10** minus **9 J/K. Despite a sample area reduction of the order of 10**6, we are able to duplicate data taken with some exfoliated graphite substrates.

Original languageEnglish (US)
Title of host publicationUnknown Host Publication Title
PublisherNorth-Holland
Pages1165-1166
Number of pages2
Editionpt 2
ISBN (Print)0444869107
StatePublished - Dec 1 1984

Publication series

Name
Numberpt 2

All Science Journal Classification (ASJC) codes

  • General Engineering

Fingerprint

Dive into the research topics of 'MICRO-CALORIMETER FOR MONOLAYER FILMS.'. Together they form a unique fingerprint.

Cite this