Miniaturized Stacked Transducer for Intravascular Sonothrombolysis With Internal-Illumination Photoacoustic Imaging Guidance and Clot Characterization

Huaiyu Wu, Yuqi Tang, Bohua Zhang, Paul Klippel, Yun Jing, Junjie Yao, Xiaoning Jiang

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

Thromboembolism in blood vessels can lead to stroke or heart attack and even sudden death unless brought under control. Sonothrombolysis enhanced by ultrasound contrast agents has shown promising outcome on effective treatment of thromboembolism. Intravascular sonothrombolysis was also reported recently with a potential for effective and safe treatment of deep thrombosis. Despite the promising treatment results, the treatment efficiency for clinical application may not be optimized due to the lack of imaging guidance and clot characterization during the thrombolysis procedure. In this paper, a miniaturized transducer was designed to have an 8-layer PZT-5A stacked with an aperture size of 1.4 × 1.4 mm2 and assembled in a customized two-lumen 10-Fr catheter for intravascular sonothrombolysis. The treatment process was monitored with internal-illumination photoacoustic tomography (II-PAT), a hybrid imaging modality that combines the rich contrast of optical absorption and the deep penetration of ultrasound detection. With intravascular light delivery using a thin optical fiber integrated with the intravascular catheter, II-PAT overcomes the penetration depth limited by strong optical attenuation of tissue. In-vitro PAT-guided sonothrombolysis experiments were carried out with synthetic blood clots embedded in tissue phantom. Clot position, shape, stiffness, and oxygenation level can be estimated by II-PAT at clinically relevant depth of ten centimeters. Our findings have demonstrated the feasibility of the proposed PAT-guided intravascular sonothrombolysis with real-time feedback during the treatment process.

Original languageEnglish (US)
Pages (from-to)2279-2288
Number of pages10
JournalIEEE Transactions on Biomedical Engineering
Volume70
Issue number8
DOIs
StatePublished - Aug 1 2023

All Science Journal Classification (ASJC) codes

  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Miniaturized Stacked Transducer for Intravascular Sonothrombolysis With Internal-Illumination Photoacoustic Imaging Guidance and Clot Characterization'. Together they form a unique fingerprint.

Cite this