Modified design of the coil probe for high field MRI

Bahram Seifi, Elena Semouchkina, Warren Perger, Gang Chea Lee, Thomas Neuberger, Michael Lanagan

Research output: Chapter in Book/Report/Conference proceedingConference contribution

2 Scopus citations

Abstract

Small RF coils for operating in Magnetic Resonance Imaging (MRI) 14T systems with Larmor frequency of 600 MHz for imaging seed assembles have been developed with the focus to improve the homogeneity of magnetic field inside the coils for obtaining higher image quality. The performed full-wave simulations have shown that uniformly wrapped coils fail to provide the desired field homogeneity, while non-uniform wrapping can be employed for achieving essentially more congruous field distributions.

Original languageEnglish (US)
Title of host publication2015 IEEE Antennas and Propagation Society International Symposium, APS 2015 - Proceedings
PublisherInstitute of Electrical and Electronics Engineers Inc.
Pages1354-1355
Number of pages2
ISBN (Electronic)9781479978151
DOIs
StatePublished - Oct 22 2015
EventIEEE Antennas and Propagation Society International Symposium, APS 2015 - Vancouver, Canada
Duration: Jul 19 2015Jul 24 2015

Publication series

NameIEEE Antennas and Propagation Society, AP-S International Symposium (Digest)
Volume2015-October
ISSN (Print)1522-3965

Other

OtherIEEE Antennas and Propagation Society International Symposium, APS 2015
Country/TerritoryCanada
CityVancouver
Period7/19/157/24/15

All Science Journal Classification (ASJC) codes

  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Modified design of the coil probe for high field MRI'. Together they form a unique fingerprint.

Cite this