Monopolar stray energy in robotic surgery

  • Douglas M. Overbey
  • , Heather Carmichael
  • , Krzysztof J. Wikiel
  • , Douglas A. Hirth
  • , Brandon C. Chapman
  • , John T. Moore
  • , Carlton C. Barnett
  • , Teresa S. Jones
  • , Thomas N. Robinson
  • , Edward L. Jones

Research output: Contribution to journalArticlepeer-review

Abstract

Introduction: Stray energy transfer from monopolar radiofrequency energy during laparoscopy can be potentially catastrophic. Robotic surgery is increasing in popularity; however, the risk of stray energy transfer during robotic surgery is unknown. The purpose of this study was to (1) quantify stray energy transfer using robotic instrumentation, (2) determine strategies to minimize the transfer of energy, and (3) compare robotic stray energy transfer to laparoscopy. Methods: In a laparoscopic trainer, a monopolar instrument (L-hook) was activated with DaVinci Si (Intuitive, Sunnyvale, CA) robotic instruments. A camera and assistant grasper were inserted to mimic a minimally invasive cholecystectomy. During activation of the L-hook, the non-electric tips of the camera and grasper were placed adjacent to simulated tissue (saline-soaked sponge). The primary outcome was change in temperature from baseline (°C) measured nearest the tip of the non-electric instrument. Results: Simulated tissue nearest the robotic grasper increased an average of 18.3 ± 5.8 °C; p < 0.001 from baseline. Tissue nearest the robotic camera tip increased (9.0 ± 2.1 °C; p < 0.001). Decreasing the power from 30 to 15 W (18.3 ± 5.8 vs. 2.6 ± 2.7 °C, p < 0.001) or using low-voltage cut mode (18.3 ± 5.8 vs. 3.1 ± 2.1 °C, p < 0.001) reduced stray energy transfer to the robotic grasper. Desiccating tissue, in contrast to open air activation, also significantly reduced stray energy transfer for the grasper (18.3 ± 5.8 vs. 0.15 ± 0.21 °C, p < 0.001) and camera (9.0 ± 2.1 vs. 0.24 ± 0.34 °C, p < 0.001). Conclusions: Stray energy transfer occurs during robotic surgery. The assistant grasper carries the highest risk for thermal injury. Similar to laparoscopy, stray energy transfer can be reduced by lowering the power setting, utilizing a low-voltage cut mode instead of coagulation mode and avoiding open air activation. These practical findings can aid surgeons performing robotic surgery to reduce injuries from stray energy.

Original languageEnglish (US)
Pages (from-to)2084-2090
Number of pages7
JournalSurgical endoscopy
Volume35
Issue number5
DOIs
StatePublished - May 2021

All Science Journal Classification (ASJC) codes

  • Surgery

Fingerprint

Dive into the research topics of 'Monopolar stray energy in robotic surgery'. Together they form a unique fingerprint.

Cite this