T1 - Multi-gradient echo with susceptibility inhomogeneity compensation (MGESIC)
T2 - Demonstration of fMRI in the olfactory cortex at 3.0 T
AU - Yang, Qing
AU - Dardzinski, Bernard J.
AU - Li, Shizhe
AU - Eslinger, Paul
AU - Smith, Michael B.
PY - 1997/3
Y1 - 1997/3
N2 - Short image acquisition times and sensitivity to magnetic susceptibility favor the use of gradient echo imaging methods in functional MRI (fMRI). However, magnetic susceptibility effects attributed to air-tissue interfaces also lead to severe signal loss in images of the large inferior frontal and lateral temporal cortices of the human brain, which renders these regions inaccessible to fMRI. The signal loss is caused by the local field gradients in the slice selection direction. A multigradient echo with magnetic susceptibility inhomogeneity compensation method (MGESIC) is proposed to overcome this problem. The MGESIC method effectively corrects the susceptibility artifacts and maintains the advantages of gradient echo methods to both BOLD sensitivity and fast image acquisition. The effectiveness of the MGESIC method is demonstrated by fMRI experimental results within the olfactory cortex.
AB - Short image acquisition times and sensitivity to magnetic susceptibility favor the use of gradient echo imaging methods in functional MRI (fMRI). However, magnetic susceptibility effects attributed to air-tissue interfaces also lead to severe signal loss in images of the large inferior frontal and lateral temporal cortices of the human brain, which renders these regions inaccessible to fMRI. The signal loss is caused by the local field gradients in the slice selection direction. A multigradient echo with magnetic susceptibility inhomogeneity compensation method (MGESIC) is proposed to overcome this problem. The MGESIC method effectively corrects the susceptibility artifacts and maintains the advantages of gradient echo methods to both BOLD sensitivity and fast image acquisition. The effectiveness of the MGESIC method is demonstrated by fMRI experimental results within the olfactory cortex.
UR - http://www.scopus.com/inward/record.url?scp=0031029807&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0031029807&partnerID=8YFLogxK
U2 - 10.1002/mrm.1910370304
DO - 10.1002/mrm.1910370304
M3 - Article
C2 - 9055220
AN - SCOPUS:0031029807
SN - 0740-3194
VL - 37
SP - 331
EP - 335
JO - Magnetic Resonance in Medicine
JF - Magnetic Resonance in Medicine
IS - 3
ER -