Abstract
The aim of the study was to compare physical and technical features of a new, improved, model of polish CPAP device (POL-CPAP2). The new model is smaller and lighter than the previous one. The new model produces higher flow and maximal working pressures 335 L/min and 19 mbar respectively. Device is also less noisy at usually applied pressures. Important novelty is a 20 minutes ramp pressure increase facilitating patient to fall asleep.
Translated title of the contribution | Evaluation of a new model for a Polish continuous positive airway pressure device (POL-CPAP 2) |
---|---|
Original language | Polish |
Pages (from-to) | 391-395 |
Number of pages | 5 |
Journal | Pneumonologia i alergologia polska : organ Polskiego Towarzystwa Ftyzjopneumonologicznego, Polskiego Towarzystwa Alergologicznego, i Instytutu Gruźlicy i Chorób Płuc |
Volume | 66 |
Issue number | 7-8 |
State | Published - 1998 |
All Science Journal Classification (ASJC) codes
- General Medicine