Abstract
We report the implementation of a swept-sine dynamic force calibration system at NIST. This system is designed to perform primary calibration of force transducers at frequencies up to 2 kHz and forces up to 2 kN. We demonstrate the use of this system to measure the dynamic sensitivity of a strain-gauge based force transducer up to 2 kHz. The current dominant source of uncertainty in the system is angular motion of the calibration assembly.
Original language | English (US) |
---|---|
State | Published - 2015 |
Event | 21st IMEKO World Congress on Measurement in Research and Industry - Prague, Czech Republic Duration: Aug 30 2015 → Sep 4 2015 |
Conference
Conference | 21st IMEKO World Congress on Measurement in Research and Industry |
---|---|
Country/Territory | Czech Republic |
City | Prague |
Period | 8/30/15 → 9/4/15 |
All Science Journal Classification (ASJC) codes
- Electrical and Electronic Engineering