Abstract
Previous works have shown that placement of a high-dielectric pad can improve image intensity in a region adjacent to the pad, or that placement of dielectric pads around a large surface of the head can improve image homogeneity on an entire plane through the head in high-field MRI. Here, experimental results show that use of high-dielectric pads around the human head can reduce the required input radiofrequency power by 50% while enhancing image signal-to-noise ratio by 20-40% throughout the cerebrum at 3 T. Thus, dielectric pads may be used to provide a relatively simple and low-cost method for improving quality and safety of MRI in a variety of applications at 3 T.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 358-362 |
| Number of pages | 5 |
| Journal | Magnetic Resonance in Medicine |
| Volume | 65 |
| Issue number | 2 |
| DOIs | |
| State | Published - Feb 2011 |
All Science Journal Classification (ASJC) codes
- Radiology Nuclear Medicine and imaging
Fingerprint
Dive into the research topics of 'Reducing SAR and enhancing cerebral signal-to-noise ratio with high permittivity padding at 3 T'. Together they form a unique fingerprint.Cite this