Abstract
In this work, we demonstrate an on-chip microfluidic flow cytometry system based on a three-dimensional (3D) hydrodynamic focusing technique, microfluidic drifting. By inducing Dean flow in a curved microfluidic channel, microfluidic drifting can be used to hydrodynamically focus cells or particles in the vertical direction and enables the 3D hydrodynamic focusing in a single-layer planar microfluidic device. Through theoretical calculation, numerical simulation, and experimental characterization, we found that the microfluidic drifting technique can be effectively applied to three-dimensionally focus microparticles with density and size equivalent to those of human CD4+ T lymphocytes. In addition, we developed a flow cytometry platform by integrating the 3D focusing device with a laser-induced fluorescence (LIF) detection system. The system was shown to provide effective high-throughput flow cytometry measurements at a rate of greater than 1700 cells s-1.
Original language | English (US) |
---|---|
Pages (from-to) | 1583-1589 |
Number of pages | 7 |
Journal | Lab on a Chip |
Volume | 9 |
Issue number | 11 |
DOIs | |
State | Published - 2009 |
All Science Journal Classification (ASJC) codes
- Bioengineering
- Biochemistry
- General Chemistry
- Biomedical Engineering