Abstract
Mechanical failure of the Björk-Shiley tilting disc prosthesis is a rare but catastrophic complication of the valve. Several previously reported cases of major strut fracture with disc embolization led to improvements in the structure of the prosthesis. A case of minor strut fracture in the improved convexoconcave model of the Björk-Shiley mitral valve prosthesis is described. The literature on mechanical failures in the Björk-Shiley mitral valve prosthesis is reviewed, and recommendations for recognition and management of this problem are made.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 65-68 |
| Number of pages | 4 |
| Journal | Annals of Thoracic Surgery |
| Volume | 40 |
| Issue number | 1 |
| DOIs | |
| State | Published - 1985 |
All Science Journal Classification (ASJC) codes
- Surgery
- Pulmonary and Respiratory Medicine
- Cardiology and Cardiovascular Medicine