Abstract
Purpose: It was aimed to identify problems and investigate processing parameters in images for treatment planning and dosimetry in Y‐90 Zevalin dose escalate trial. Method and Materials: 5 mCi of In‐111 Zevalin was imaged (0, 4, 24, 72 and 144 hours) on Siemens eCam dual head gamma camera with MEGP collimators. Y‐90 patient imaging was performed with 28 mCi of Y‐90 labeled Zevalin. At 30 minutes post Zevalin injection, a whole body scan was performed on the same camera using a 194 cm bed length (5 cm/min, total 48 min), 256 × 1024 × 16 matrix without zooming. Both anterior and posterior images were transferred an IDL workstation for data analysis. Time activity curves were fitted with various functions for comparison. MIRD scheme was used to estimate patient dose. Results: In‐ 111 images were processed with attenuation compensation. Quality control in acquisition and processing were important for consistency in dosimetry calculation. Mass correction was obtained using CT scan. Most of organ dose were under 2,000 rads if Y‐90 activity was below 32 mCi. Kidney dose was sensitive to RoI definition and location of the background region. Conclusion: An energy window (75keV/50%) is adequate for characteristic X‐ray and Bremsstrahlung imaging. Y‐90 imaging depends on scatter characteristic of tissue and does not provide direct biodistribution of Yttrium‐90. In‐111 imaging simulates the distribution of Y‐90 for dosimetry when quality control and consistency are followed.
Original language | English (US) |
---|---|
Pages (from-to) | 2674 |
Number of pages | 1 |
Journal | Medical Physics |
Volume | 35 |
Issue number | 6 |
DOIs | |
State | Published - Jun 2008 |
All Science Journal Classification (ASJC) codes
- Biophysics
- Radiology Nuclear Medicine and imaging