The ability of two chlorine dioxide chemistries to inactivate human papillomavirus-contaminated endocavitary ultrasound probes and nasendoscopes

Craig Meyers, Janice Milici, Richard Robison

Research output: Contribution to journalArticlepeer-review

10 Scopus citations

Abstract

Sexual transmission is the most common pathway for the spread of Human papillomavirus (HPV). However, the potential for iatrogenic HPV infections is also real. Even though cleared by the Food and Drug Administration and recommended by the World Federation for Ultrasound in Medicine and Biology, several disinfectants including glutaraldehyde and o-phthalaldehyde have shown a lack of efficacy for inactivating HPV. Other methods such as ultraviolet C and concentrated hydrogen peroxide have been shown highly effective at inactivating infectious HPV. In this study, two chlorine dioxide systems are also shown to be highly efficacious at inactivating HPV. An important difference in these present studies is that as opposed to testing in suspension or using a carrier, we dried the infectious virus directly onto endocavitary ultrasound probes and nasendoscopes, therefore, validating a more realistic system to demonstrate disinfectant efficacy.

Original languageEnglish (US)
Pages (from-to)1298-1302
Number of pages5
JournalJournal of Medical Virology
Volume92
Issue number8
DOIs
StatePublished - Aug 1 2020

All Science Journal Classification (ASJC) codes

  • Virology
  • Infectious Diseases

Fingerprint

Dive into the research topics of 'The ability of two chlorine dioxide chemistries to inactivate human papillomavirus-contaminated endocavitary ultrasound probes and nasendoscopes'. Together they form a unique fingerprint.