T1 - The Accelerated Transcatheter Heart Valve Testing Environment
T2 - Loading, Motion, and Fluid Dynamics
AU - Ponnaluri, Sailahari V.
AU - Sacks, Michael S.
AU - Manning, Keefe B.
N1 - Publisher Copyright:
Copyright © 2023 by ASME.
PY - 2023/3/1
Y1 - 2023/3/1
N2 - Transcatheter aortic valve replacements (TAVRs) are an increasingly common treatment for aortic valve disease due to their minimally invasive delivery. As TAVR designs require thinner leaflets to facilitate catheter-based delivery, they experience greater leaflet operational stresses and potentially greater durability issues than conventional surgical valves. Yet, our understanding of TAVR durability remains largely unexplored. Currently, preclinical TAVR durability is evaluated within an ISO:5840 compliant accelerated wear tester (AWT) up to a required 200 × 106 cycles, corresponding to approximately five years in vivo. While AWTs use high cycle frequencies (10-20 Hz) to achieve realistic timeframes, the resulting valve loading behaviors and fluid dynamics are not representative of the in vivo environment and thus may not accurately predict failure mechanisms. Despite the importance of fatigue and failure predictions for replacement heart valves, surprisingly, little quantitative information exists on the dynamic AWT environment. To better understand this environment, we examined frequency and diastolic period effects for the first time using high-speed enface imaging and particle image velocimetry to quantify valve motion and flow, respectively, using a Durapulse™ AWT at frequencies of 10, 15, and 20 Hz. Regardless of operating condition, no waveform achieved a physiologically relevant transvalvular loading pressure, despite having an ISO compliant geometric orifice area (GOA) and waveform. General fluid mechanics were consistent with in vivo but the AWT geometry developed secondary flow structures, which could impact mechanical loading. Therefore, the nonphysiologic loading and variability induced by changes in operating condition must be carefully regulated to ensure physiologically relevant fatigue.
AB - Transcatheter aortic valve replacements (TAVRs) are an increasingly common treatment for aortic valve disease due to their minimally invasive delivery. As TAVR designs require thinner leaflets to facilitate catheter-based delivery, they experience greater leaflet operational stresses and potentially greater durability issues than conventional surgical valves. Yet, our understanding of TAVR durability remains largely unexplored. Currently, preclinical TAVR durability is evaluated within an ISO:5840 compliant accelerated wear tester (AWT) up to a required 200 × 106 cycles, corresponding to approximately five years in vivo. While AWTs use high cycle frequencies (10-20 Hz) to achieve realistic timeframes, the resulting valve loading behaviors and fluid dynamics are not representative of the in vivo environment and thus may not accurately predict failure mechanisms. Despite the importance of fatigue and failure predictions for replacement heart valves, surprisingly, little quantitative information exists on the dynamic AWT environment. To better understand this environment, we examined frequency and diastolic period effects for the first time using high-speed enface imaging and particle image velocimetry to quantify valve motion and flow, respectively, using a Durapulse™ AWT at frequencies of 10, 15, and 20 Hz. Regardless of operating condition, no waveform achieved a physiologically relevant transvalvular loading pressure, despite having an ISO compliant geometric orifice area (GOA) and waveform. General fluid mechanics were consistent with in vivo but the AWT geometry developed secondary flow structures, which could impact mechanical loading. Therefore, the nonphysiologic loading and variability induced by changes in operating condition must be carefully regulated to ensure physiologically relevant fatigue.
UR - http://www.scopus.com/inward/record.url?scp=85143644577&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85143644577&partnerID=8YFLogxK
U2 - 10.1115/1.4056291
DO - 10.1115/1.4056291
M3 - Article
C2 - 36416296
AN - SCOPUS:85143644577
SN - 0148-0731
VL - 145
JO - Journal of Biomechanical Engineering
JF - Journal of Biomechanical Engineering
IS - 3
ER -