TY - JOUR
T1 - The Denali trial
T2 - An interim analysis of a prospective, multicenter study of the Denali retrievable inferior vena cava filter
AU - Stavropoulos, S. William
AU - Sing, Ronald F.
AU - Elmasri, Fakhir
AU - Silver, Mitchell J.
AU - Powell, Alex
AU - Lynch, Frank C.
AU - Aal, Ahmed Kamel Abdel
AU - Lansky, Alexandra J.
AU - Settlage, Richard A.
AU - Muhs, Bart E.
N1 - Publisher Copyright:
© 2014 SIR.
PY - 2014/10/1
Y1 - 2014/10/1
N2 - Purpose To assess safety and effectiveness of a nitinol retrievable inferior vena cava (IVC) filter in patients who require caval interruption to protect against pulmonary embolism (PE).Materials and Methods Two hundred patients with temporary indications for an IVC filter were enrolled in this prospective, multicenter clinical study. Patients undergoing filter implantation were to be followed for 2 years or for 30 days after filter retrieval. At the time of the present interim report, all 200 patients had been enrolled in the study, and 160 had undergone a retrieval attempt or been followed to 6 months with their filter in place. Primary study endpoints included technical and clinical success of filter placement and retrieval. Patients were also evaluated for recurrent PE, new or worsening deep vein thrombosis, and filter migration, fracture, penetration, and tilt.Results Clinical success of placement was achieved in 94.5% of patients (172 of 182), with a one-sided lower limit of the 95% confidence interval of 90.1%. Technical success rate of filter placement was 99.5%. Technical success rate of retrieval was 97.3%; 108 filters were retrieved in 111 attempts. In two cases, the filter apex could not be engaged with a snare, and one device was engaged but could not be removed. Filter retrievals occurred at a mean indwell time of 165 days (range, 5-632 d). There were no instances of filter fracture, migration, or tilt greater than 15° at the time of retrieval or 6-month follow-up.Conclusions In this interim report, the nitinol retrievable IVC filter provided protection against pulmonary embolism, and the device could be retrieved with a low rate of complications.
AB - Purpose To assess safety and effectiveness of a nitinol retrievable inferior vena cava (IVC) filter in patients who require caval interruption to protect against pulmonary embolism (PE).Materials and Methods Two hundred patients with temporary indications for an IVC filter were enrolled in this prospective, multicenter clinical study. Patients undergoing filter implantation were to be followed for 2 years or for 30 days after filter retrieval. At the time of the present interim report, all 200 patients had been enrolled in the study, and 160 had undergone a retrieval attempt or been followed to 6 months with their filter in place. Primary study endpoints included technical and clinical success of filter placement and retrieval. Patients were also evaluated for recurrent PE, new or worsening deep vein thrombosis, and filter migration, fracture, penetration, and tilt.Results Clinical success of placement was achieved in 94.5% of patients (172 of 182), with a one-sided lower limit of the 95% confidence interval of 90.1%. Technical success rate of filter placement was 99.5%. Technical success rate of retrieval was 97.3%; 108 filters were retrieved in 111 attempts. In two cases, the filter apex could not be engaged with a snare, and one device was engaged but could not be removed. Filter retrievals occurred at a mean indwell time of 165 days (range, 5-632 d). There were no instances of filter fracture, migration, or tilt greater than 15° at the time of retrieval or 6-month follow-up.Conclusions In this interim report, the nitinol retrievable IVC filter provided protection against pulmonary embolism, and the device could be retrieved with a low rate of complications.
UR - http://www.scopus.com/inward/record.url?scp=84908158464&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84908158464&partnerID=8YFLogxK
U2 - 10.1016/j.jvir.2014.07.001
DO - 10.1016/j.jvir.2014.07.001
M3 - Article
C2 - 25066514
AN - SCOPUS:84908158464
SN - 1051-0443
VL - 25
SP - 1497-1505.e1
JO - Journal of Vascular and Interventional Radiology
JF - Journal of Vascular and Interventional Radiology
IS - 10
ER -