The feasibility of nitric oxide delivery with high frequency jet ventilation

Yaseen Arabi, Anand Kumar, Kenneth Wood, Anne Flaten

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

A 27-year-old female with acute monocytic leukaemia (M5) developed acute respiratory distress syndrome (ARDS). Because of refractory hypoxaemia and severe barotrauma during conventional mechanical ventilation, the patient was switched to high frequency jet ventilation (HFJV) as a salvage therapy. Her refractory hypoxaemia improved temporarily but worsened again. Approval of the Institutional Review Board and the Food and Drug Administration was obtained to use nitric oxide (NO) with HFJV on a compassionate basis considering the grave situation. The NO delivery system was connected to the secondary flow circuit of the HFJV immediately after the humidifier. We measured the concentration of NO at least every hour by inserting a 7-Fr catheter connected to a McNeill analyzer into the endotracheal tube. Despite improvement of oxygenation, the patient's respiratory status deteriorated further and she died. In this case we were able to achieve reliable and constant levels of NO. The purpose of this report is to discuss the technical aspects and pitfalls of this method.

Original languageEnglish (US)
Pages (from-to)673-677
Number of pages5
JournalRespirology
Volume10
Issue number5
DOIs
StatePublished - Nov 2005

All Science Journal Classification (ASJC) codes

  • Pulmonary and Respiratory Medicine

Fingerprint

Dive into the research topics of 'The feasibility of nitric oxide delivery with high frequency jet ventilation'. Together they form a unique fingerprint.

Cite this