T1 - Traceable calibration and demonstration of a portable dynamic force transfer standard
AU - Vlajic, Nicholas
AU - Chijioke, Ako
N1 - Publisher Copyright:
© 2017 Not subject to copyright in the USA. Contribution of NIST.
PY - 2017/7/18
Y1 - 2017/7/18
N2 - In general, the dynamic sensitivity of a force transducer depends upon the mechanical system in which it is used. This dependence serves as motivation to develop a dynamic force transfer standard, which can be used to calibrate an application transducer in situ. In this work, we SI-traceably calibrate a hand-held force transducer, namely an impact hammer, by using a mass suspended from a thin line which is cut to produce a known dynamic force in the form of a step function. We show that this instrument is a promising candidate as a transfer standard, since its dynamic response has small variance between different users. This calibrated transfer standard is then used to calibrate a secondary force transducer in an example application setting. The combined standard uncertainty (k = 2) in the calibration of the transfer standard was determined to be 2.1% or less, up to a bandwidth of 5 kHz. The combined standard uncertainty (k = 2) in the performed transfer calibration was less than 4%, up to 3 kHz. An advantage of the transfer calibration framework presented here, is that the transfer standard can be used to transfer SI-traceable calibrations without the use of any SI-traceable voltage metrology instrumentation.
AB - In general, the dynamic sensitivity of a force transducer depends upon the mechanical system in which it is used. This dependence serves as motivation to develop a dynamic force transfer standard, which can be used to calibrate an application transducer in situ. In this work, we SI-traceably calibrate a hand-held force transducer, namely an impact hammer, by using a mass suspended from a thin line which is cut to produce a known dynamic force in the form of a step function. We show that this instrument is a promising candidate as a transfer standard, since its dynamic response has small variance between different users. This calibrated transfer standard is then used to calibrate a secondary force transducer in an example application setting. The combined standard uncertainty (k = 2) in the calibration of the transfer standard was determined to be 2.1% or less, up to a bandwidth of 5 kHz. The combined standard uncertainty (k = 2) in the performed transfer calibration was less than 4%, up to 3 kHz. An advantage of the transfer calibration framework presented here, is that the transfer standard can be used to transfer SI-traceable calibrations without the use of any SI-traceable voltage metrology instrumentation.
UR - http://www.scopus.com/inward/record.url?scp=85026367822&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85026367822&partnerID=8YFLogxK
U2 - 10.1088/1681-7575/aa75da
DO - 10.1088/1681-7575/aa75da
M3 - Article
AN - SCOPUS:85026367822
SN - 0026-1394
VL - 54
SP - S83-S98
JO - Metrologia
JF - Metrologia
IS - 4
ER -