Treadmill exercise in calves implanted with the AbioCor replacement heart

Benjamin B. Chiang, George M. Pantalos, Patricia Cerrito, Karla Stevens, Karen Lott, Laura Lott, Steven Etoch, Laman A. Gray, Robert D. Dowling

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

The aim of this study was to evaluate the function of the AbioCor artificial heart implanted in calves that underwent treadmill exercise testing and to investigate the cardiovascular and metabolic responses during exercise. Six calves were implanted with the AbioCor. One week after surgery, exercise training was undertaken and animals were put on a treadmill at a speed of 0.5-1.0 mph, with the AbioCor operating at the maximum heart rate of 150 beats per minute. During exercise, the left and right atrial pressures were significantly increased (p < 0.05). Oxygen consumption was significantly elevated (p < 0.05). Lactate levels did not significantly increase (p > 0.05). The index of metabolic adequacy remained higher than 2.0, indicating that there was no substantial transition to anaerobic metabolism during exercise. There was an increased extraction of oxygen during exercise to meet the increased energy demands. Our study demonstrated that animals implanted with the AbioCor heart were able to maintain aerobic metabolism during low levels of exercise testing.

Original languageEnglish (US)
Pages (from-to)589-593
Number of pages5
JournalASAIO Journal
Volume49
Issue number5
DOIs
StatePublished - Sep 2003

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Bioengineering
  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Treadmill exercise in calves implanted with the AbioCor replacement heart'. Together they form a unique fingerprint.

Cite this