T1 - Validation of the NITI endoluminal compression anastomosis ring (EndoCAR) device and comparison to the traditional circular stapled colorectal anastomosis in a porcine model
AU - Stewart, David
AU - Hunt, Steven
AU - Pierce, Richard
AU - Mao, Dongli
AU - Frisella, Margaret
AU - Cook, Kathryn
AU - Starcher, Barry
AU - Fleshman, James
N1 - Copyright:
Copyright 2008 Elsevier B.V., All rights reserved.
PY - 2007/12
Y1 - 2007/12
N2 - The purpose of this study was to determine whether the characteristics of compression anastomoses created by a new device are comparable to existing stapler technology. A total of 18 pigs were studied, and each served as its own control using a 27-mm compression device and a 29-mm stapler. Anastomoses were randomized to proximal and distal positions along the rectum and were separated by 10 cm. Six nonsurvival pigs were sacrificed at zero time to failure test the anastomoses. Twelve pigs were sacrificed at 14 days. Failure pressures, circumferences, and radiographic leak rates were determined. Anastomotic tissue was processed for matrix metalloproteinase, collagen, and elastin levels. The compression anastomoses had higher mean failure pressures than stapled anastomoses at zero time (103 vs 29.9 mm Hg). At 2 weeks, there was no difference between failure pressures (256 vs 250 mm Hg). During burst testing, 3 of the compression anastomoses failed at the anastomosis at 2 weeks, whereas none of the stapled anastomoses failed. The mean anastomotic circumference of the compression anastomoses was narrower than the stapled anastomoses (9.63 vs 11.25 cm, P = .001). There were no clinical leaks or radiographic leaks by barium enema at 2 weeks. There was no difference between matrix metalloproteinase, collagen, or elastin content based on tissue analysis. There were dense adhesions to 7 of 12 (58.3%) of the stapled anastomoses, whereas only 1 of 12 (8.3%) of the compression anastomoses had flimsy adhesions. A new compression anastomosis technique using a nickel-titanium alloy may be an advance in technology by reducing leaks and eliminating foreign material in the anastomosis.
AB - The purpose of this study was to determine whether the characteristics of compression anastomoses created by a new device are comparable to existing stapler technology. A total of 18 pigs were studied, and each served as its own control using a 27-mm compression device and a 29-mm stapler. Anastomoses were randomized to proximal and distal positions along the rectum and were separated by 10 cm. Six nonsurvival pigs were sacrificed at zero time to failure test the anastomoses. Twelve pigs were sacrificed at 14 days. Failure pressures, circumferences, and radiographic leak rates were determined. Anastomotic tissue was processed for matrix metalloproteinase, collagen, and elastin levels. The compression anastomoses had higher mean failure pressures than stapled anastomoses at zero time (103 vs 29.9 mm Hg). At 2 weeks, there was no difference between failure pressures (256 vs 250 mm Hg). During burst testing, 3 of the compression anastomoses failed at the anastomosis at 2 weeks, whereas none of the stapled anastomoses failed. The mean anastomotic circumference of the compression anastomoses was narrower than the stapled anastomoses (9.63 vs 11.25 cm, P = .001). There were no clinical leaks or radiographic leaks by barium enema at 2 weeks. There was no difference between matrix metalloproteinase, collagen, or elastin content based on tissue analysis. There were dense adhesions to 7 of 12 (58.3%) of the stapled anastomoses, whereas only 1 of 12 (8.3%) of the compression anastomoses had flimsy adhesions. A new compression anastomosis technique using a nickel-titanium alloy may be an advance in technology by reducing leaks and eliminating foreign material in the anastomosis.
UR - http://www.scopus.com/inward/record.url?scp=37449000460&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=37449000460&partnerID=8YFLogxK
U2 - 10.1177/1553350607312241
DO - 10.1177/1553350607312241
M3 - Article
C2 - 18178912
AN - SCOPUS:37449000460
SN - 1553-3506
VL - 14
SP - 252
EP - 260
JO - Surgical Innovation
JF - Surgical Innovation
IS - 4
ER -